510(k) Data Aggregation

    K Number
    K191484

    Validate with FDA (Live)

    Date Cleared
    2019-07-30

    (56 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Renuvion®/J-Plasma® Precise Handpiece is intended to be used with compatible electrosurgical generators for the delivery of helium plasma for cutting, coagulation and ablation of soft tissue during open and laparoscopic surgical procedures.

    The Renuvion®/J-Plasma® Precise Handpiece is compatible with the Electrosurgical Generators BVX-200H and BVX-200P manufactured by Apyx Medical.

    Device Description

    The Renuvion®/J-Plasma® Precise Handpiece is a sterile, single use electrosurgical accessory intended to be used in conjunction with compatible electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating, and ablating soft tissue. The compatible Generators operate at an adjustable power of up to 40W (expressed as 0-100% where 100% is 40W) and provide an adjustable helium gas flow of 1-5 LPM. The Renuvion/J-Plasma Precise Handpiece is available with a retractable cutting tip blade for excising tissue. The blade tip serves as an electrode to generate helium plasma.

    The blade can be extended in 2mm increments from 0 to 10mm from the distal tip of the shaft via a ratcheted slider and retracted via a release latch. The device is activated via pressing the purple activation trigger. The handpiece is to be used with compatible electrosurgical generators, BVX-200H or BVX-200P, manufactured by Apyx Medical.

    The Renuvion®/J-Plasma® Precise Handpiece is available in a 33 cm length for use in laparoscopic or open surgical procedures. The 33 cm is a standard length for laparoscopic instruments and are compatible with standard 5 mm trocars.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from Apyx Medical Corporation for their Renuvion®/J-Plasma® Precise Handpiece (Model # BVX-330 BPS). This submission describes a modification of an already cleared device and aims to demonstrate substantial equivalence to previously marketed predicate devices. Therefore, the "acceptance criteria" and "device performance" described are focused on proving this equivalence through various tests rather than establishing initial clinical efficacy with novel endpoints.

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria / Test ObjectiveReported Device Performance (Result)
    Mechanical Performance & Functionality: Verify Mechanical functionality of the Renuvion®/J-Plasma® Precise Handpiece.Mechanical functionality requirements met.
    Tissue Effect (Plasma Activation): Verify that the tissue effects are similar between the subject device and the Primary Predicate device (K183610 & K151325).Tissue effects are equivalent between the subject device and Primary Predicate device.
    Tissue Effect (Monopolar Coagulation): Verify that the tissue effects are similar between the subject device and the Reference device (K183610 & K170188).Tissue effects are equivalent between the subject device and Reference device.
    Electrical Safety: Compliance with IEC 60601-1 and IEC 60601-2-2 standards.Determined to be in conformance with these standards.
    Electromagnetic Compatibility (EMC): Compliance with IEC 60601-1-2 standard.Determined to be in conformance with this standard.
    Biocompatibility: Evaluation in accordance with FDA guidance "Use of International Standard ISO 10993-1" for cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, and material mediated pyrogen/pyrogenicity.Biocompatibility has been established per ISO 10993 guidelines for the device's classification (external communicating device with indirect blood path contact < 24 hours). Biocompatibility verification testing was satisfactorily conducted.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes (e.g., number of units tested) for the individual bench tests. The tests were performed as "bench testing," meaning they were conducted in a laboratory setting, not on human or animal subjects. The data provenance is internal to Apyx Medical Corporation and based on their design and development procedures and risk assessment in accordance with ISO 14971. No geographical provenance (country of origin for data) is specified beyond the company's location in Clearwater, Florida, USA. The tests are retrospective in the sense that they are comparing a modified device to previously cleared predicates, rather than a de novo clinical investigation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. For bench testing related to mechanical functionality, electrical safety, EMC, and biocompatibility, "experts" in the traditional sense of medical specialists establishing ground truth (like radiologists reading images) are not typically involved. Instead, the ground truth is defined by established engineering standards, material science principles, and biological safety guidelines. For tissue effect tests, conclusions are drawn from bench experiments and comparison to predicate devices, where the "ground truth" would be the observed tissue response under controlled conditions, compared to the known response of the predicate. The document doesn't detail the personnel or their qualifications who interpreted these tissue effects.

    4. Adjudication Method for the Test Set:

    An adjudication method (like 2+1 or 3+1 expert review) is not applicable here. These methods are common for clinical imaging studies where there's subjectivity in interpretation. For the bench tests described (mechanical, electrical, EMC, biocompatibility, and comparing tissue effects), the results are typically objectively measured and evaluated against predefined thresholds or comparative equivalence, not through expert consensus on subjective findings.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not performed, nor is it applicable to this submission. MRMC studies are used to evaluate the diagnostic performance of a device (often AI-based) in a clinical setting with multiple human readers and cases, typically to assess how the device aids human interpretation. This submission is for an electrosurgical handpiece, and the performance assessment is based on physical and biological characteristics, not diagnostic interpretation by human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    A standalone performance study is not applicable here. This device is a surgical instrument, not an algorithm or AI system. Its function is directly performed by the device itself, often with a human operator, but its primary performance endpoints are physical effects (cutting, coagulation) and safety, not an algorithmic output that could be independently evaluated.

    7. The Type of Ground Truth Used:

    The ground truth used for these tests is generally based on:

    • Engineering Specifications and Standards: For mechanical performance, electrical safety, and EMC, the ground truth is defined by established industry and international standards (e.g., IEC 60601 series) and the device's own design specifications.
    • Predicate Device Performance: For tissue effects, the ground truth for "equivalence" is the documented and cleared performance of the primary predicate device (for J-Plasma mode) and the reference device (for Monopolar Coagulation mode). The results of the subject device are compared to these established performances.
    • Biological Safety Guidelines: For biocompatibility, the ground truth is established by international standards like ISO 10993 guidelines, which define acceptable biological responses for device materials.

    8. The Sample Size for the Training Set:

    This information is not applicable because this is a medical device, not an AI/ML product that requires a training set for algorithm development. The "design controls process" and "verification and validation" mentioned refer to engineering and quality assurance processes for medical devices, not to machine learning training.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable as there is no training set for an AI/ML algorithm involved in this device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183610

    Validate with FDA (Live)

    Date Cleared
    2019-03-12

    (76 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Renuvion®/J-Plasma® Precise Handpiece is intended to be used with compatible J-Plasma electrosurgical generators for the delivery of helium plasma for cutting, coagulation of soft tissue during open and laparoscopic surgical procedures.

    The Renuvion®/J-Plasma® Precise Open Handpiece is intended to be used with compatible J-Plasma electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Renuvion®/J-Plasma® Precise Open® Handpieces are compatible with electrosurgical Generators BVX-200H, and BVX-200P.

    Device Description

    The Renuvion/J-Plasma Handpieces are sterile, single use electrosurgical accessories intended to be used in conjunction with compatible electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating soft tissue. The compatible Generators operate at an adjustable power of up to 40W (expressed as 0-100% is 40W) and provide an adjustable helium gas flow of 1-5 LPM. The Renuvion/J-Plasma Handpieces are available with a retractable cutting blade tip for excising tissue. The blade tip serves as an electrode to generate helium plasma.

    The Renuvion/J-Plasma Precise devices are available in 15cm, 27cm and 33cm lengths for use in laparoscopic or open surgical procedures. The 27cm are standard lengths for laparoscopic instruments and are compatible with standard 5mm trocars.

    The Renuvion/J-Plasma Precise Open devices are available in 44mm and 150mm for used in open surgical procedures.

    AI/ML Overview

    This document describes the Renuvion/J-Plasma Precise Handpiece and Renuvion/J-Plasma Precise Open Handpiece, intended for use with compatible electrosurgical generators for soft tissue cutting, coagulation, and ablation during open and laparoscopic surgical procedures.

    Here's an analysis of the acceptance criteria and the supporting studies:

    1. Table of acceptance criteria and the reported device performance:

      Test / Acceptance CriteriaReported Device Performance
      Mechanical & Electrical Verification and Functionality
      Pneumatic TestingVerified
      Electrical Continuity TestVerified
      Cable Flexure / Tension TestVerified
      Connector Dielectric Strength TestVerified
      Mold Stress TestVerified
      Free-fall Drop TestVerified
      Insertion/Extraction Integrity (Reduced insertion/extraction force to plug/unplug)Verified (The document specifically notes a reduction in insertion/extraction force compared to predicate devices, indicating this was a design change tested and met)
      Usability EvaluationProtocol executed to verify and validate usability "as it relates to safety." (Satisfactorily conducted)
      Biocompatibility Testing (for external communicating device with indirect blood path, < 24 hours)Conducted per FDA Blue Book Memo #G95-1 and ISO 10993-1. Included Cytotoxicity, Sensitization, Irritation, Systemic toxicity, Hemolysis, and Material Mediated Pyrogen. Verification testing for changed patient-contacting components (flow tube coupler and O-ring) was satisfactorily conducted.
    2. Sample size used for the test set and the data provenance:

      The document does not explicitly state the sample sizes for each mechanical, electrical, usability, or biocompatibility test. It mentions "Performance testing" was performed, indicating that multiple units were likely tested for each criterion. The data provenance is internal testing performed by the manufacturer, Bovie Medical Corporation. The tests are prospective as they were performed on the new device models to demonstrate their safety and efficacy.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      This information is not provided. The tests described are primarily engineering, electrical, and biological evaluations, typically not requiring medical experts for "ground truth" establishment in the same way clinical studies would. The "usability evaluation" would have involved users, but their qualifications are not specified beyond being relevant to using the "J-Plasma Pistol Grip Handpiece."

    4. Adjudication method for the test set:

      This information is not provided and is not typically relevant for the type of engineering and biocompatibility testing described. The tests would likely have pass/fail criteria based on predefined specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      No, an MRMC comparative effectiveness study was not done. This device is an electrosurgical handpiece, not an AI diagnostic or assistance tool, so such a study would not be applicable.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      No, this is an electrosurgical handpiece, a physical medical device, not a software algorithm. Therefore, "standalone" algorithm performance is not applicable.

    7. The type of ground truth used:

      The ground truth for the various tests was based on:

      • Engineering Specifications: For mechanical and electrical tests (e.g., connector dielectric strength, cable flexure, integrity of insertion/extraction force).
      • Regulatory Standards: For biocompatibility testing (ISO 10993 guidelines, FDA Blue Book Memorandum #G95-1).
      • Safety and Efficacy Requirements: For usability evaluation, ensuring the device is safe and performs as intended.
    8. The sample size for the training set:

      This concept is not applicable as this is a physical medical device, not a machine learning algorithm requiring a training set.

    9. How the ground truth for the training set was established:

      This concept is not applicable as this is a physical medical device, not a machine learning algorithm requiring a training set.